Ethylene Oxide (EO) Sterilizers
 
															 
															 
															 
															 
															 
															Especially designed for sterilization with plain Ethylene Oxide or with an explosion-proof mixture of Ethylene Oxide and an inert gas.
Chamber
									Doors
									Control Panel
									Heating System
									Information Record
									Loading System
									Ex Proof
									Accesories
									Models
							Chamber
					Entirely made of stainless steel 316L and calculated according to the Asme code. The inner areas have sanitary finish. The whole chamber is thermally insulated.
									Doors
					Vertical or horizontal airtight sliding doors with silicone gasket. Double-door sterilizers fitted with a safety system which prevents both doors from opening at the same time. 
									Control Panel
					Microprocessed command with industrial PLC and HMI touch screen, Title 21 CFR Part 11 compliant, which guarantees data traceability. There are different access levels through passwords to set the different parameters for each cycle.
									Heating System
					Designed to work with steam coming from a central line. If this is not a possibility, the equipment can be delivered with its own built-in electrically heated steam generator.
									Information Record
					An alphanumeric printer prints all process data and variables.
									Loading System
					Composed of loading trolleys and transfer trolleys to place and transfer the materials to be processed inside the autoclave. Every loading system is analyzed and designed based on the different kind of materials to be processed, this being a key element in the design of an autoclave. 
									Ex Proof
					When working with plain Ethylene Oxide the equipment will be built with all the necessary safety measures required to avoid an explosion. Equipment complies with the Ex Proof safety standard.
									Accesories
					Scrubber. It neutralizes the Ethylene Oxide after degasification stage. 
Scales. Ethylene Oxide cross control and dosing system which allows for dosing of the gas for each cycle and control of excessive consumption or gas leak.
									Scales. Ethylene Oxide cross control and dosing system which allows for dosing of the gas for each cycle and control of excessive consumption or gas leak.
Models
					| Models | Capacity (L) | 
|---|---|
| 662a | 544 L | 
| 882 | 768 L | 
| 992 | 1032 L | 
| 1120 | 1200 L | 
| 1140 | 2400 L | 
*We manufacture customized chambers for capacity greater than 30,000 L.
 
								 
															 
															